SPECT imaging of 131I (364 keV): Importance of collimation

L. P. Clarke, C. B. Saw, L. K. Leong, Aldo N Serafini

Research output: Contribution to journalArticle

7 Citations (Scopus)

Abstract

A low sensitivity medium energy collimator (LSMEC) designed with thick septa and long bore (theoretical leakage < 3% 364 keV), was evaluated using 131I for a SPECT system operated in both planar and tomographic imaging modes. The collimator was designed to minimize the influence of photon penetration on spatial resolution, in particular the resolution index FWTM. Overall spatial resolution for the planar imaging mode at 10 cm from the collimator face was found to be 11.6 mm FWHM and 21.6 mm FWTM. The corresponding transverse plane and slice thickness resolution was of the order of 17 mm FWHM and 31 mm FWTM, for a radius of rotation of 16 cm. A SPECT resolution phantom was imaged. Two quadrants of cold rods were well resolved, with rod dimensions of 16 and 12.7 mm respectively, the resolution being comparable to that obtained using 99Tc(m) (140 keV) and a low-energy high-resolution collimator. NEMA sensitivity obtained was 75 cpm/μCi 131I. The resolution measurements obtained suggest that this collimator should be useful for SPECT imaging with 131I in either radioimmunoimaging or radioimmunotherapy.

Original languageEnglish
Pages (from-to)41-47
Number of pages7
JournalNuclear Medicine Communications
Volume6
Issue number1
StatePublished - Jan 1 1985

Fingerprint

Single-Photon Emission-Computed Tomography
Radioimmunodetection
Radioimmunotherapy
Photons

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

SPECT imaging of 131I (364 keV) : Importance of collimation. / Clarke, L. P.; Saw, C. B.; Leong, L. K.; Serafini, Aldo N.

In: Nuclear Medicine Communications, Vol. 6, No. 1, 01.01.1985, p. 41-47.

Research output: Contribution to journalArticle

Clarke, L. P. ; Saw, C. B. ; Leong, L. K. ; Serafini, Aldo N. / SPECT imaging of 131I (364 keV) : Importance of collimation. In: Nuclear Medicine Communications. 1985 ; Vol. 6, No. 1. pp. 41-47.
@article{d1ea222424a140fa879a9f158c9d166b,
title = "SPECT imaging of 131I (364 keV): Importance of collimation",
abstract = "A low sensitivity medium energy collimator (LSMEC) designed with thick septa and long bore (theoretical leakage < 3{\%} 364 keV), was evaluated using 131I for a SPECT system operated in both planar and tomographic imaging modes. The collimator was designed to minimize the influence of photon penetration on spatial resolution, in particular the resolution index FWTM. Overall spatial resolution for the planar imaging mode at 10 cm from the collimator face was found to be 11.6 mm FWHM and 21.6 mm FWTM. The corresponding transverse plane and slice thickness resolution was of the order of 17 mm FWHM and 31 mm FWTM, for a radius of rotation of 16 cm. A SPECT resolution phantom was imaged. Two quadrants of cold rods were well resolved, with rod dimensions of 16 and 12.7 mm respectively, the resolution being comparable to that obtained using 99Tc(m) (140 keV) and a low-energy high-resolution collimator. NEMA sensitivity obtained was 75 cpm/μCi 131I. The resolution measurements obtained suggest that this collimator should be useful for SPECT imaging with 131I in either radioimmunoimaging or radioimmunotherapy.",
author = "Clarke, {L. P.} and Saw, {C. B.} and Leong, {L. K.} and Serafini, {Aldo N}",
year = "1985",
month = "1",
day = "1",
language = "English",
volume = "6",
pages = "41--47",
journal = "Nuclear Medicine Communications",
issn = "0143-3636",
publisher = "Lippincott Williams and Wilkins",
number = "1",

}

TY - JOUR

T1 - SPECT imaging of 131I (364 keV)

T2 - Importance of collimation

AU - Clarke, L. P.

AU - Saw, C. B.

AU - Leong, L. K.

AU - Serafini, Aldo N

PY - 1985/1/1

Y1 - 1985/1/1

N2 - A low sensitivity medium energy collimator (LSMEC) designed with thick septa and long bore (theoretical leakage < 3% 364 keV), was evaluated using 131I for a SPECT system operated in both planar and tomographic imaging modes. The collimator was designed to minimize the influence of photon penetration on spatial resolution, in particular the resolution index FWTM. Overall spatial resolution for the planar imaging mode at 10 cm from the collimator face was found to be 11.6 mm FWHM and 21.6 mm FWTM. The corresponding transverse plane and slice thickness resolution was of the order of 17 mm FWHM and 31 mm FWTM, for a radius of rotation of 16 cm. A SPECT resolution phantom was imaged. Two quadrants of cold rods were well resolved, with rod dimensions of 16 and 12.7 mm respectively, the resolution being comparable to that obtained using 99Tc(m) (140 keV) and a low-energy high-resolution collimator. NEMA sensitivity obtained was 75 cpm/μCi 131I. The resolution measurements obtained suggest that this collimator should be useful for SPECT imaging with 131I in either radioimmunoimaging or radioimmunotherapy.

AB - A low sensitivity medium energy collimator (LSMEC) designed with thick septa and long bore (theoretical leakage < 3% 364 keV), was evaluated using 131I for a SPECT system operated in both planar and tomographic imaging modes. The collimator was designed to minimize the influence of photon penetration on spatial resolution, in particular the resolution index FWTM. Overall spatial resolution for the planar imaging mode at 10 cm from the collimator face was found to be 11.6 mm FWHM and 21.6 mm FWTM. The corresponding transverse plane and slice thickness resolution was of the order of 17 mm FWHM and 31 mm FWTM, for a radius of rotation of 16 cm. A SPECT resolution phantom was imaged. Two quadrants of cold rods were well resolved, with rod dimensions of 16 and 12.7 mm respectively, the resolution being comparable to that obtained using 99Tc(m) (140 keV) and a low-energy high-resolution collimator. NEMA sensitivity obtained was 75 cpm/μCi 131I. The resolution measurements obtained suggest that this collimator should be useful for SPECT imaging with 131I in either radioimmunoimaging or radioimmunotherapy.

UR - http://www.scopus.com/inward/record.url?scp=0021889854&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0021889854&partnerID=8YFLogxK

M3 - Article

C2 - 3879947

AN - SCOPUS:0021889854

VL - 6

SP - 41

EP - 47

JO - Nuclear Medicine Communications

JF - Nuclear Medicine Communications

SN - 0143-3636

IS - 1

ER -