Optical cochlear implants: Evaluation of surgical approach and laser parameters in cats

Suhrud M Rajguru, Agnella Izzo Matic, Alan M. Robinson, Andrew J. Fishman, Laura E. Moreno, Allison Bradley, Irena Vujanovic, Joe Breen, Jonathon D. Wells, Mark Bendett, Claus Peter Richter

Research output: Contribution to journalArticle

63 Citations (Scopus)

Abstract

Previous research has shown that neural stimulation with infrared radiation (IR) is spatially selective and illustrated the potential of IR in stimulating auditory neurons. The present work demonstrates the application of a miniaturized pulsed IR stimulator for chronic implantation in cats, quantifies its efficacy, and short-term safety in stimulating auditory neurons. IR stimulation of the neurons was achieved using an optical fiber inserted through a cochleostomy drilled in the basal turn of the cat cochlea and was characterized by measuring compound action potentials (CAPs). Neurons were stimulated with IR at various pulse durations, radiant exposures, and pulse repetition rates. Pulse durations as short as 50 μs were successful in evoking CAPs in normal as well as deafened cochleae. Continual stimulation was provided at 200 pulses per second, at 200 mW per pulse, and 100 μs pulse duration. Stable CAP amplitudes were observed for up to 10 h of continual IR stimulation. Combined with histological data, the results suggest that pulsed IR stimulation does not lead to detectable acute tissue damage and validate the stimulation parameters that can be used in future chronic implants based on pulsed IR.

Original languageEnglish
Pages (from-to)102-111
Number of pages10
JournalHearing Research
Volume269
Issue number1-2
DOIs
StatePublished - Oct 1 2010
Externally publishedYes

Fingerprint

Cochlear Implants
Cats
Lasers
Radiation
Action Potentials
Neurons
Cochlea
Optical Fibers
Heart Rate
Safety
Research

ASJC Scopus subject areas

  • Sensory Systems

Cite this

Rajguru, S. M., Matic, A. I., Robinson, A. M., Fishman, A. J., Moreno, L. E., Bradley, A., ... Richter, C. P. (2010). Optical cochlear implants: Evaluation of surgical approach and laser parameters in cats. Hearing Research, 269(1-2), 102-111. https://doi.org/10.1016/j.heares.2010.06.021

Optical cochlear implants : Evaluation of surgical approach and laser parameters in cats. / Rajguru, Suhrud M; Matic, Agnella Izzo; Robinson, Alan M.; Fishman, Andrew J.; Moreno, Laura E.; Bradley, Allison; Vujanovic, Irena; Breen, Joe; Wells, Jonathon D.; Bendett, Mark; Richter, Claus Peter.

In: Hearing Research, Vol. 269, No. 1-2, 01.10.2010, p. 102-111.

Research output: Contribution to journalArticle

Rajguru, SM, Matic, AI, Robinson, AM, Fishman, AJ, Moreno, LE, Bradley, A, Vujanovic, I, Breen, J, Wells, JD, Bendett, M & Richter, CP 2010, 'Optical cochlear implants: Evaluation of surgical approach and laser parameters in cats', Hearing Research, vol. 269, no. 1-2, pp. 102-111. https://doi.org/10.1016/j.heares.2010.06.021
Rajguru, Suhrud M ; Matic, Agnella Izzo ; Robinson, Alan M. ; Fishman, Andrew J. ; Moreno, Laura E. ; Bradley, Allison ; Vujanovic, Irena ; Breen, Joe ; Wells, Jonathon D. ; Bendett, Mark ; Richter, Claus Peter. / Optical cochlear implants : Evaluation of surgical approach and laser parameters in cats. In: Hearing Research. 2010 ; Vol. 269, No. 1-2. pp. 102-111.
@article{3a5898ea79554cc39d2a9e0731fe55ad,
title = "Optical cochlear implants: Evaluation of surgical approach and laser parameters in cats",
abstract = "Previous research has shown that neural stimulation with infrared radiation (IR) is spatially selective and illustrated the potential of IR in stimulating auditory neurons. The present work demonstrates the application of a miniaturized pulsed IR stimulator for chronic implantation in cats, quantifies its efficacy, and short-term safety in stimulating auditory neurons. IR stimulation of the neurons was achieved using an optical fiber inserted through a cochleostomy drilled in the basal turn of the cat cochlea and was characterized by measuring compound action potentials (CAPs). Neurons were stimulated with IR at various pulse durations, radiant exposures, and pulse repetition rates. Pulse durations as short as 50 μs were successful in evoking CAPs in normal as well as deafened cochleae. Continual stimulation was provided at 200 pulses per second, at 200 mW per pulse, and 100 μs pulse duration. Stable CAP amplitudes were observed for up to 10 h of continual IR stimulation. Combined with histological data, the results suggest that pulsed IR stimulation does not lead to detectable acute tissue damage and validate the stimulation parameters that can be used in future chronic implants based on pulsed IR.",
author = "Rajguru, {Suhrud M} and Matic, {Agnella Izzo} and Robinson, {Alan M.} and Fishman, {Andrew J.} and Moreno, {Laura E.} and Allison Bradley and Irena Vujanovic and Joe Breen and Wells, {Jonathon D.} and Mark Bendett and Richter, {Claus Peter}",
year = "2010",
month = "10",
day = "1",
doi = "10.1016/j.heares.2010.06.021",
language = "English",
volume = "269",
pages = "102--111",
journal = "Hearing Research",
issn = "0378-5955",
publisher = "Elsevier",
number = "1-2",

}

TY - JOUR

T1 - Optical cochlear implants

T2 - Evaluation of surgical approach and laser parameters in cats

AU - Rajguru, Suhrud M

AU - Matic, Agnella Izzo

AU - Robinson, Alan M.

AU - Fishman, Andrew J.

AU - Moreno, Laura E.

AU - Bradley, Allison

AU - Vujanovic, Irena

AU - Breen, Joe

AU - Wells, Jonathon D.

AU - Bendett, Mark

AU - Richter, Claus Peter

PY - 2010/10/1

Y1 - 2010/10/1

N2 - Previous research has shown that neural stimulation with infrared radiation (IR) is spatially selective and illustrated the potential of IR in stimulating auditory neurons. The present work demonstrates the application of a miniaturized pulsed IR stimulator for chronic implantation in cats, quantifies its efficacy, and short-term safety in stimulating auditory neurons. IR stimulation of the neurons was achieved using an optical fiber inserted through a cochleostomy drilled in the basal turn of the cat cochlea and was characterized by measuring compound action potentials (CAPs). Neurons were stimulated with IR at various pulse durations, radiant exposures, and pulse repetition rates. Pulse durations as short as 50 μs were successful in evoking CAPs in normal as well as deafened cochleae. Continual stimulation was provided at 200 pulses per second, at 200 mW per pulse, and 100 μs pulse duration. Stable CAP amplitudes were observed for up to 10 h of continual IR stimulation. Combined with histological data, the results suggest that pulsed IR stimulation does not lead to detectable acute tissue damage and validate the stimulation parameters that can be used in future chronic implants based on pulsed IR.

AB - Previous research has shown that neural stimulation with infrared radiation (IR) is spatially selective and illustrated the potential of IR in stimulating auditory neurons. The present work demonstrates the application of a miniaturized pulsed IR stimulator for chronic implantation in cats, quantifies its efficacy, and short-term safety in stimulating auditory neurons. IR stimulation of the neurons was achieved using an optical fiber inserted through a cochleostomy drilled in the basal turn of the cat cochlea and was characterized by measuring compound action potentials (CAPs). Neurons were stimulated with IR at various pulse durations, radiant exposures, and pulse repetition rates. Pulse durations as short as 50 μs were successful in evoking CAPs in normal as well as deafened cochleae. Continual stimulation was provided at 200 pulses per second, at 200 mW per pulse, and 100 μs pulse duration. Stable CAP amplitudes were observed for up to 10 h of continual IR stimulation. Combined with histological data, the results suggest that pulsed IR stimulation does not lead to detectable acute tissue damage and validate the stimulation parameters that can be used in future chronic implants based on pulsed IR.

UR - http://www.scopus.com/inward/record.url?scp=77956345012&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77956345012&partnerID=8YFLogxK

U2 - 10.1016/j.heares.2010.06.021

DO - 10.1016/j.heares.2010.06.021

M3 - Article

C2 - 20603207

AN - SCOPUS:77956345012

VL - 269

SP - 102

EP - 111

JO - Hearing Research

JF - Hearing Research

SN - 0378-5955

IS - 1-2

ER -