Normative retardation data corrected for the corneal polarization axis with scanning laser polarimetry

David Greenfield, Robert W. Knighton, William J Feuer, Joyce C. Schiffman

Research output: Contribution to journalArticle

14 Citations (Scopus)

Abstract

BACKGROUND AND OBJECTIVE: To evaluate the distribution of retinal nerve fiber layer thickness measurements in normal eyes corrected for corneal polarization axis (CPA). METHODS: Complete ocular examination, standard automated perimetry, peripapillary and macular scanning laser polarimetry imaging, and CPA measurements were performed in normal eyes. A noninvasive device mounted on a slit lamp that incorporated two crossed linear polarizers and an optical retarder to measure the slow axis of corneal birefringence was constructed. One eye per patient was enrolled. Exclusion criteria consisted of visual acuity of less than 20/40, a refractive error exceeding ± 5.0 diopter sphere, 2.0 diopter cylinder, or both, previous intraocular surgery, or ocular disease except cataract. According to scanning laser polarimetry, peripapillary retardation parameters within 90% normal limits were recorded as "normal," parameters outside 95% normal limits were recorded as "abnormal," and parameters between 90% and 95% limits were recorded as "borderline." RESULTS: Fifty-one eyes of 51 subjects (14 men, 37 women) were enrolled (mean age, 51 ± 17 years). Prediction limits (± 2 standard deviation outside of regression line) were calculated for 14 peripapillary retardation parameters. The prediction limits of the average thickness of the retinal nerve fiber layer (μm) were reduced 35% by correction of the CPA; for ellipse average (pm), the limits were reduced 33%. Sixteen of 51 normal eyes (31%) had at least one abnormal retardation parameter, and among these 16 eyes there was a bimodal distribution of CPA values (0° to 20° nasally downward, 60° to 80° nasally downward). CONCLUSIONS: Correction for CPA reduces the variance of normal retardation measurements. Normal eyes with a CPA of 0° to 20° nasally downward and 60° to 80° nasally downward were characterized by "abnormal" summary retardation parameters as interpreted by scanning laser polarimetry.

Original languageEnglish
Pages (from-to)165-171
Number of pages7
JournalOphthalmic Surgery Lasers and Imaging
Volume34
Issue number2
StatePublished - Mar 1 2003

Fingerprint

Scanning Laser Polarimetry
Nerve Fibers
Birefringence
Visual Field Tests
Refractive Errors
Eye Diseases
Cataract
Visual Acuity
Equipment and Supplies

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Normative retardation data corrected for the corneal polarization axis with scanning laser polarimetry. / Greenfield, David; Knighton, Robert W.; Feuer, William J; Schiffman, Joyce C.

In: Ophthalmic Surgery Lasers and Imaging, Vol. 34, No. 2, 01.03.2003, p. 165-171.

Research output: Contribution to journalArticle

Greenfield, David ; Knighton, Robert W. ; Feuer, William J ; Schiffman, Joyce C. / Normative retardation data corrected for the corneal polarization axis with scanning laser polarimetry. In: Ophthalmic Surgery Lasers and Imaging. 2003 ; Vol. 34, No. 2. pp. 165-171.
@article{2b36cc59d1b04693a3cde0e6a7a54803,
title = "Normative retardation data corrected for the corneal polarization axis with scanning laser polarimetry",
abstract = "BACKGROUND AND OBJECTIVE: To evaluate the distribution of retinal nerve fiber layer thickness measurements in normal eyes corrected for corneal polarization axis (CPA). METHODS: Complete ocular examination, standard automated perimetry, peripapillary and macular scanning laser polarimetry imaging, and CPA measurements were performed in normal eyes. A noninvasive device mounted on a slit lamp that incorporated two crossed linear polarizers and an optical retarder to measure the slow axis of corneal birefringence was constructed. One eye per patient was enrolled. Exclusion criteria consisted of visual acuity of less than 20/40, a refractive error exceeding ± 5.0 diopter sphere, 2.0 diopter cylinder, or both, previous intraocular surgery, or ocular disease except cataract. According to scanning laser polarimetry, peripapillary retardation parameters within 90{\%} normal limits were recorded as {"}normal,{"} parameters outside 95{\%} normal limits were recorded as {"}abnormal,{"} and parameters between 90{\%} and 95{\%} limits were recorded as {"}borderline.{"} RESULTS: Fifty-one eyes of 51 subjects (14 men, 37 women) were enrolled (mean age, 51 ± 17 years). Prediction limits (± 2 standard deviation outside of regression line) were calculated for 14 peripapillary retardation parameters. The prediction limits of the average thickness of the retinal nerve fiber layer (μm) were reduced 35{\%} by correction of the CPA; for ellipse average (pm), the limits were reduced 33{\%}. Sixteen of 51 normal eyes (31{\%}) had at least one abnormal retardation parameter, and among these 16 eyes there was a bimodal distribution of CPA values (0° to 20° nasally downward, 60° to 80° nasally downward). CONCLUSIONS: Correction for CPA reduces the variance of normal retardation measurements. Normal eyes with a CPA of 0° to 20° nasally downward and 60° to 80° nasally downward were characterized by {"}abnormal{"} summary retardation parameters as interpreted by scanning laser polarimetry.",
author = "David Greenfield and Knighton, {Robert W.} and Feuer, {William J} and Schiffman, {Joyce C.}",
year = "2003",
month = "3",
day = "1",
language = "English",
volume = "34",
pages = "165--171",
journal = "Ophthalmic Surgery Lasers and Imaging Retina",
issn = "2325-8160",
publisher = "Slack Incorporated",
number = "2",

}

TY - JOUR

T1 - Normative retardation data corrected for the corneal polarization axis with scanning laser polarimetry

AU - Greenfield, David

AU - Knighton, Robert W.

AU - Feuer, William J

AU - Schiffman, Joyce C.

PY - 2003/3/1

Y1 - 2003/3/1

N2 - BACKGROUND AND OBJECTIVE: To evaluate the distribution of retinal nerve fiber layer thickness measurements in normal eyes corrected for corneal polarization axis (CPA). METHODS: Complete ocular examination, standard automated perimetry, peripapillary and macular scanning laser polarimetry imaging, and CPA measurements were performed in normal eyes. A noninvasive device mounted on a slit lamp that incorporated two crossed linear polarizers and an optical retarder to measure the slow axis of corneal birefringence was constructed. One eye per patient was enrolled. Exclusion criteria consisted of visual acuity of less than 20/40, a refractive error exceeding ± 5.0 diopter sphere, 2.0 diopter cylinder, or both, previous intraocular surgery, or ocular disease except cataract. According to scanning laser polarimetry, peripapillary retardation parameters within 90% normal limits were recorded as "normal," parameters outside 95% normal limits were recorded as "abnormal," and parameters between 90% and 95% limits were recorded as "borderline." RESULTS: Fifty-one eyes of 51 subjects (14 men, 37 women) were enrolled (mean age, 51 ± 17 years). Prediction limits (± 2 standard deviation outside of regression line) were calculated for 14 peripapillary retardation parameters. The prediction limits of the average thickness of the retinal nerve fiber layer (μm) were reduced 35% by correction of the CPA; for ellipse average (pm), the limits were reduced 33%. Sixteen of 51 normal eyes (31%) had at least one abnormal retardation parameter, and among these 16 eyes there was a bimodal distribution of CPA values (0° to 20° nasally downward, 60° to 80° nasally downward). CONCLUSIONS: Correction for CPA reduces the variance of normal retardation measurements. Normal eyes with a CPA of 0° to 20° nasally downward and 60° to 80° nasally downward were characterized by "abnormal" summary retardation parameters as interpreted by scanning laser polarimetry.

AB - BACKGROUND AND OBJECTIVE: To evaluate the distribution of retinal nerve fiber layer thickness measurements in normal eyes corrected for corneal polarization axis (CPA). METHODS: Complete ocular examination, standard automated perimetry, peripapillary and macular scanning laser polarimetry imaging, and CPA measurements were performed in normal eyes. A noninvasive device mounted on a slit lamp that incorporated two crossed linear polarizers and an optical retarder to measure the slow axis of corneal birefringence was constructed. One eye per patient was enrolled. Exclusion criteria consisted of visual acuity of less than 20/40, a refractive error exceeding ± 5.0 diopter sphere, 2.0 diopter cylinder, or both, previous intraocular surgery, or ocular disease except cataract. According to scanning laser polarimetry, peripapillary retardation parameters within 90% normal limits were recorded as "normal," parameters outside 95% normal limits were recorded as "abnormal," and parameters between 90% and 95% limits were recorded as "borderline." RESULTS: Fifty-one eyes of 51 subjects (14 men, 37 women) were enrolled (mean age, 51 ± 17 years). Prediction limits (± 2 standard deviation outside of regression line) were calculated for 14 peripapillary retardation parameters. The prediction limits of the average thickness of the retinal nerve fiber layer (μm) were reduced 35% by correction of the CPA; for ellipse average (pm), the limits were reduced 33%. Sixteen of 51 normal eyes (31%) had at least one abnormal retardation parameter, and among these 16 eyes there was a bimodal distribution of CPA values (0° to 20° nasally downward, 60° to 80° nasally downward). CONCLUSIONS: Correction for CPA reduces the variance of normal retardation measurements. Normal eyes with a CPA of 0° to 20° nasally downward and 60° to 80° nasally downward were characterized by "abnormal" summary retardation parameters as interpreted by scanning laser polarimetry.

UR - http://www.scopus.com/inward/record.url?scp=0037358653&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037358653&partnerID=8YFLogxK

M3 - Article

C2 - 12665235

AN - SCOPUS:0037358653

VL - 34

SP - 165

EP - 171

JO - Ophthalmic Surgery Lasers and Imaging Retina

JF - Ophthalmic Surgery Lasers and Imaging Retina

SN - 2325-8160

IS - 2

ER -