Abstract
Various portable monitors have been used to quantify physical activity but most rely on detecting limb movement with a sensor rather than measuring muscle activity. Our first goal was to design and validate a portable system for recording surface electromyographic activity (EMG) from eight muscles over 24 h. The modular system includes: (1) preamplifiers that filter and amplify signals; (2) a preprocessor unit for further filtering and amplification, signal offset and power supply modification; (3) a data-logger for analog-to-digital conversion; a flash memory card for data storage and (4) a rechargeable battery. The equipment samples EMG at 1000 Hz, has a resolution of 2.6 μV and records signals up to 10 mV. The built-in analog filters create a bandwidth appropriate for surface EMG. Our second aim was to test the system biologically by recording EMG from able-bodied and spinal cord injured participants. Modifications were made to electrodes for remote preamplifier placement, and to the battery connection after pilot testing. Thereafter, 31 consecutive 24-h EMG recordings were successful. Both the engineering and biological validation of this system establishes it as a valuable tool for measuring physical activity from different muscles in real-world environments whether individuals have an intact or damaged nervous system.
Original language | English (US) |
---|---|
Pages (from-to) | 81-91 |
Number of pages | 11 |
Journal | Biomedizinische Technik |
Volume | 64 |
Issue number | 1 |
DOIs | |
State | Published - Feb 1 2019 |
Keywords
- physical activity
- portable recordings
- spinal cord injury
ASJC Scopus subject areas
- Biomedical Engineering