Graft patency verification in coronary artery bypass grafting: Principles and clinical applications of transit time flow measurement

G. D'Ancona, H. L. Karamanoukian, M. Ricci, J. Bergsland, Tomas Salerno

Research output: Contribution to journalArticle

27 Citations (Scopus)

Abstract

The increasing popularity of beating-heart coronary surgery has raised concerns and doubts about the quality of the coronary anastomoses performed. Intraoperative graft patency verification methods are not commonly used after coronary surgery and, most of the cardiac surgeons, rely on the simple clinical signs (electrocardiogram tracings and hemodynamic stability) to make a diagnosis of coronary graft occlusion. New transit time ultrasound based methods for graft-patency verification have been adopted in many centers during beating-heart and traditional bypass grafting. Although the results are very encouraging, correct interpretation of the flow findings may prove difficult if specific rules are not properly followed. Flow curves, pulsatility index, and flow values should always be considered simultaneously before revising a coronary graft. Measurements should also be always performed with and without a proximal coronary snare. This article summarizes the main features of flowmetry and provides some technical pitfalls and suggestions to achieve an adequate intraoperative flow measurement adopting the transit time method.

Original languageEnglish
Pages (from-to)725-731
Number of pages7
JournalAngiology
Volume51
Issue number9
StatePublished - Oct 3 2000
Externally publishedYes

Fingerprint

Coronary Artery Bypass
Transplants
Thoracic Surgery
Rheology
Coronary Occlusion
Electrocardiography
Hemodynamics

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Graft patency verification in coronary artery bypass grafting : Principles and clinical applications of transit time flow measurement. / D'Ancona, G.; Karamanoukian, H. L.; Ricci, M.; Bergsland, J.; Salerno, Tomas.

In: Angiology, Vol. 51, No. 9, 03.10.2000, p. 725-731.

Research output: Contribution to journalArticle

D'Ancona, G. ; Karamanoukian, H. L. ; Ricci, M. ; Bergsland, J. ; Salerno, Tomas. / Graft patency verification in coronary artery bypass grafting : Principles and clinical applications of transit time flow measurement. In: Angiology. 2000 ; Vol. 51, No. 9. pp. 725-731.
@article{856363ef6949476e92b85332d4bfbd53,
title = "Graft patency verification in coronary artery bypass grafting: Principles and clinical applications of transit time flow measurement",
abstract = "The increasing popularity of beating-heart coronary surgery has raised concerns and doubts about the quality of the coronary anastomoses performed. Intraoperative graft patency verification methods are not commonly used after coronary surgery and, most of the cardiac surgeons, rely on the simple clinical signs (electrocardiogram tracings and hemodynamic stability) to make a diagnosis of coronary graft occlusion. New transit time ultrasound based methods for graft-patency verification have been adopted in many centers during beating-heart and traditional bypass grafting. Although the results are very encouraging, correct interpretation of the flow findings may prove difficult if specific rules are not properly followed. Flow curves, pulsatility index, and flow values should always be considered simultaneously before revising a coronary graft. Measurements should also be always performed with and without a proximal coronary snare. This article summarizes the main features of flowmetry and provides some technical pitfalls and suggestions to achieve an adequate intraoperative flow measurement adopting the transit time method.",
author = "G. D'Ancona and Karamanoukian, {H. L.} and M. Ricci and J. Bergsland and Tomas Salerno",
year = "2000",
month = "10",
day = "3",
language = "English",
volume = "51",
pages = "725--731",
journal = "Angiology",
issn = "0003-3197",
publisher = "SAGE Publications Inc.",
number = "9",

}

TY - JOUR

T1 - Graft patency verification in coronary artery bypass grafting

T2 - Principles and clinical applications of transit time flow measurement

AU - D'Ancona, G.

AU - Karamanoukian, H. L.

AU - Ricci, M.

AU - Bergsland, J.

AU - Salerno, Tomas

PY - 2000/10/3

Y1 - 2000/10/3

N2 - The increasing popularity of beating-heart coronary surgery has raised concerns and doubts about the quality of the coronary anastomoses performed. Intraoperative graft patency verification methods are not commonly used after coronary surgery and, most of the cardiac surgeons, rely on the simple clinical signs (electrocardiogram tracings and hemodynamic stability) to make a diagnosis of coronary graft occlusion. New transit time ultrasound based methods for graft-patency verification have been adopted in many centers during beating-heart and traditional bypass grafting. Although the results are very encouraging, correct interpretation of the flow findings may prove difficult if specific rules are not properly followed. Flow curves, pulsatility index, and flow values should always be considered simultaneously before revising a coronary graft. Measurements should also be always performed with and without a proximal coronary snare. This article summarizes the main features of flowmetry and provides some technical pitfalls and suggestions to achieve an adequate intraoperative flow measurement adopting the transit time method.

AB - The increasing popularity of beating-heart coronary surgery has raised concerns and doubts about the quality of the coronary anastomoses performed. Intraoperative graft patency verification methods are not commonly used after coronary surgery and, most of the cardiac surgeons, rely on the simple clinical signs (electrocardiogram tracings and hemodynamic stability) to make a diagnosis of coronary graft occlusion. New transit time ultrasound based methods for graft-patency verification have been adopted in many centers during beating-heart and traditional bypass grafting. Although the results are very encouraging, correct interpretation of the flow findings may prove difficult if specific rules are not properly followed. Flow curves, pulsatility index, and flow values should always be considered simultaneously before revising a coronary graft. Measurements should also be always performed with and without a proximal coronary snare. This article summarizes the main features of flowmetry and provides some technical pitfalls and suggestions to achieve an adequate intraoperative flow measurement adopting the transit time method.

UR - http://www.scopus.com/inward/record.url?scp=0033829208&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0033829208&partnerID=8YFLogxK

M3 - Article

C2 - 10999613

AN - SCOPUS:0033829208

VL - 51

SP - 725

EP - 731

JO - Angiology

JF - Angiology

SN - 0003-3197

IS - 9