Determination of picomolar iron in seawater by double Mg(OH)2 precipitation isotope dilution high-resolution ICPMS

Research output: Contribution to journalArticle

41 Scopus citations

Abstract

A low-blank pre-concentration procedure is described for the analysis of picomolar iron (Fe) in seawater by isotope dilution high-resolution inductively coupled plasma mass-spectrometry (HR-ICPMS). The procedure uses a two-step Mg(OH)2 co-precipitation procedure to extract Fe from a 50 ml seawater sample into a 100 μl 4% nitric acid (HNO3) solution followed by HR-ICPMS measurement. The high pre-concentration ratio (∼ 500:1) achieved by the procedure minimizes the Fe blank due to ICPMS instrumental Fe background and results in a detection limit of ∼ 2 pM and a precision of ∼ 4% at the 50 pM Fe level. The measurement of a low-Fe seawater sample spiked with gravimetric Fe standard shows that the method can clearly distinguish 0.01 nM Fe from 0.02 nM Fe in seawater with high accuracy. The method is demonstrated by the analysis of dissolved Fe in the equatorial Pacific Ocean.

Original languageEnglish (US)
Pages (from-to)370-381
Number of pages12
JournalMarine Chemistry
Volume103
Issue number3-4
DOIs
StatePublished - Jan 16 2007
Externally publishedYes

Keywords

  • Analysis
  • Iron
  • Seawater
  • Trace metal

ASJC Scopus subject areas

  • Oceanography
  • Chemistry(all)
  • Environmental Chemistry
  • Water Science and Technology

Fingerprint Dive into the research topics of 'Determination of picomolar iron in seawater by double Mg(OH)<sub>2</sub> precipitation isotope dilution high-resolution ICPMS'. Together they form a unique fingerprint.

  • Cite this