Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants

Abhishek Prasad, Qing Shan Xue, Viswanath Sankar, Toshikazu Nishida, Gerry Shaw, Wolfgang J. Streit, Justin C. Sanchez

Research output: Contribution to journalArticle

143 Citations (Scopus)

Abstract

For nearly 55 years, tungsten microwires have been widely used in neurophysiological experiments in animal models to chronically record neuronal activity. While tungsten microwires initially provide stable recordings, their inability to reliably record high-quality neural signals for tens of years has limited their efficacy for neuroprosthetic applications in humans. Comprehensive understanding of the mechanisms of electrode performance and failure is necessary for developing next generation neural interfaces for humans. In this study, we evaluated the abiotic (electrophysiology, impedance, electrode morphology) and biotic (microglial reactivity, blood-brain barrier disruption, biochemical markers of axonal injury) effects of 16-channel, 50 μm diameter, polyimide insulated tungsten microwires array for implant durations that ranged from acute to up to 9 months in 25 rats. Daily electrode impedance spectroscopy, electrophysiological recordings, blood and cerebrospinal fluid (CSF) withdrawals, and histopathological analysis were performed to study the time-varying effects of chronic electrode implantation. Structural changes at the electrode recording site were observed as early as within 2-3 h of electrode insertion. Abiotic analysis indicated the first 2-3 weeks following surgery was the most dynamic period in the chronic electrode lifetime as there were greater variations in the electrode impedance, functional electrode performance, and the structural changes occurring at the electrode recording tips. Electrode recording site deterioration continued for the long-term chronic animals as insulation damage occurred and recording surface became more recessed over time. In general, electrode impedance and functional performance had smaller daily variations combined with reduced electrode recording site changes during the chronic phase. Histopathological studies were focused largely on characterizing microglial cell responses to electrode implantation. We found that activated microglia were present near the electrode tracks in all non-acute animals studied, thus indicating presence of a neuroinflammatory response regardless of post-implantation survival times and electrode performance. Conversely, dystrophic microglia detectable as fragmented cells were found almost exclusively in acute animals surviving only few hours after implantation. While there was no consistent relationship between microglial cell responses and electrode performance, we noticed co-occurrence of high ferritin expression, intraparenchymal bleeding, and microglial degeneration suggesting presence of excessive oxidative stress via Fenton chemistry. Biochemical analysis indicated that these electrodes always caused a persistent release of axonal injury biomarkers even several months after implantation suggesting persistent tissue damage. Our study suggests that mechanisms of electrode failure are multi-factorial involving both abiotic and biotic parameters. Since these failure modes occur concurrently and cannot be isolated from one another, the lack of consistent relationship between electrode performance and microglial responses in our results suggest that one or more of the abiotic factors were equally responsible for degradation in electrode performance over long periods of time.

Original languageEnglish
Article number056015
JournalJournal of Neural Engineering
Volume9
Issue number5
DOIs
StatePublished - Oct 1 2012

Fingerprint

Tungsten
Failure modes
Electrodes
Animals
Electric Impedance
Microglia
Biomarkers
Electrophysiology
Cerebrospinal fluid
Dielectric Spectroscopy

ASJC Scopus subject areas

  • Biomedical Engineering
  • Cellular and Molecular Neuroscience

Cite this

Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants. / Prasad, Abhishek; Xue, Qing Shan; Sankar, Viswanath; Nishida, Toshikazu; Shaw, Gerry; Streit, Wolfgang J.; Sanchez, Justin C.

In: Journal of Neural Engineering, Vol. 9, No. 5, 056015, 01.10.2012.

Research output: Contribution to journalArticle

Prasad, Abhishek ; Xue, Qing Shan ; Sankar, Viswanath ; Nishida, Toshikazu ; Shaw, Gerry ; Streit, Wolfgang J. ; Sanchez, Justin C. / Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants. In: Journal of Neural Engineering. 2012 ; Vol. 9, No. 5.
@article{53075957478348a18dd828c5f692113e,
title = "Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants",
abstract = "For nearly 55 years, tungsten microwires have been widely used in neurophysiological experiments in animal models to chronically record neuronal activity. While tungsten microwires initially provide stable recordings, their inability to reliably record high-quality neural signals for tens of years has limited their efficacy for neuroprosthetic applications in humans. Comprehensive understanding of the mechanisms of electrode performance and failure is necessary for developing next generation neural interfaces for humans. In this study, we evaluated the abiotic (electrophysiology, impedance, electrode morphology) and biotic (microglial reactivity, blood-brain barrier disruption, biochemical markers of axonal injury) effects of 16-channel, 50 μm diameter, polyimide insulated tungsten microwires array for implant durations that ranged from acute to up to 9 months in 25 rats. Daily electrode impedance spectroscopy, electrophysiological recordings, blood and cerebrospinal fluid (CSF) withdrawals, and histopathological analysis were performed to study the time-varying effects of chronic electrode implantation. Structural changes at the electrode recording site were observed as early as within 2-3 h of electrode insertion. Abiotic analysis indicated the first 2-3 weeks following surgery was the most dynamic period in the chronic electrode lifetime as there were greater variations in the electrode impedance, functional electrode performance, and the structural changes occurring at the electrode recording tips. Electrode recording site deterioration continued for the long-term chronic animals as insulation damage occurred and recording surface became more recessed over time. In general, electrode impedance and functional performance had smaller daily variations combined with reduced electrode recording site changes during the chronic phase. Histopathological studies were focused largely on characterizing microglial cell responses to electrode implantation. We found that activated microglia were present near the electrode tracks in all non-acute animals studied, thus indicating presence of a neuroinflammatory response regardless of post-implantation survival times and electrode performance. Conversely, dystrophic microglia detectable as fragmented cells were found almost exclusively in acute animals surviving only few hours after implantation. While there was no consistent relationship between microglial cell responses and electrode performance, we noticed co-occurrence of high ferritin expression, intraparenchymal bleeding, and microglial degeneration suggesting presence of excessive oxidative stress via Fenton chemistry. Biochemical analysis indicated that these electrodes always caused a persistent release of axonal injury biomarkers even several months after implantation suggesting persistent tissue damage. Our study suggests that mechanisms of electrode failure are multi-factorial involving both abiotic and biotic parameters. Since these failure modes occur concurrently and cannot be isolated from one another, the lack of consistent relationship between electrode performance and microglial responses in our results suggest that one or more of the abiotic factors were equally responsible for degradation in electrode performance over long periods of time.",
author = "Abhishek Prasad and Xue, {Qing Shan} and Viswanath Sankar and Toshikazu Nishida and Gerry Shaw and Streit, {Wolfgang J.} and Sanchez, {Justin C.}",
year = "2012",
month = "10",
day = "1",
doi = "10.1088/1741-2560/9/5/056015",
language = "English",
volume = "9",
journal = "Journal of Neural Engineering",
issn = "1741-2560",
publisher = "IOP Publishing Ltd.",
number = "5",

}

TY - JOUR

T1 - Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants

AU - Prasad, Abhishek

AU - Xue, Qing Shan

AU - Sankar, Viswanath

AU - Nishida, Toshikazu

AU - Shaw, Gerry

AU - Streit, Wolfgang J.

AU - Sanchez, Justin C.

PY - 2012/10/1

Y1 - 2012/10/1

N2 - For nearly 55 years, tungsten microwires have been widely used in neurophysiological experiments in animal models to chronically record neuronal activity. While tungsten microwires initially provide stable recordings, their inability to reliably record high-quality neural signals for tens of years has limited their efficacy for neuroprosthetic applications in humans. Comprehensive understanding of the mechanisms of electrode performance and failure is necessary for developing next generation neural interfaces for humans. In this study, we evaluated the abiotic (electrophysiology, impedance, electrode morphology) and biotic (microglial reactivity, blood-brain barrier disruption, biochemical markers of axonal injury) effects of 16-channel, 50 μm diameter, polyimide insulated tungsten microwires array for implant durations that ranged from acute to up to 9 months in 25 rats. Daily electrode impedance spectroscopy, electrophysiological recordings, blood and cerebrospinal fluid (CSF) withdrawals, and histopathological analysis were performed to study the time-varying effects of chronic electrode implantation. Structural changes at the electrode recording site were observed as early as within 2-3 h of electrode insertion. Abiotic analysis indicated the first 2-3 weeks following surgery was the most dynamic period in the chronic electrode lifetime as there were greater variations in the electrode impedance, functional electrode performance, and the structural changes occurring at the electrode recording tips. Electrode recording site deterioration continued for the long-term chronic animals as insulation damage occurred and recording surface became more recessed over time. In general, electrode impedance and functional performance had smaller daily variations combined with reduced electrode recording site changes during the chronic phase. Histopathological studies were focused largely on characterizing microglial cell responses to electrode implantation. We found that activated microglia were present near the electrode tracks in all non-acute animals studied, thus indicating presence of a neuroinflammatory response regardless of post-implantation survival times and electrode performance. Conversely, dystrophic microglia detectable as fragmented cells were found almost exclusively in acute animals surviving only few hours after implantation. While there was no consistent relationship between microglial cell responses and electrode performance, we noticed co-occurrence of high ferritin expression, intraparenchymal bleeding, and microglial degeneration suggesting presence of excessive oxidative stress via Fenton chemistry. Biochemical analysis indicated that these electrodes always caused a persistent release of axonal injury biomarkers even several months after implantation suggesting persistent tissue damage. Our study suggests that mechanisms of electrode failure are multi-factorial involving both abiotic and biotic parameters. Since these failure modes occur concurrently and cannot be isolated from one another, the lack of consistent relationship between electrode performance and microglial responses in our results suggest that one or more of the abiotic factors were equally responsible for degradation in electrode performance over long periods of time.

AB - For nearly 55 years, tungsten microwires have been widely used in neurophysiological experiments in animal models to chronically record neuronal activity. While tungsten microwires initially provide stable recordings, their inability to reliably record high-quality neural signals for tens of years has limited their efficacy for neuroprosthetic applications in humans. Comprehensive understanding of the mechanisms of electrode performance and failure is necessary for developing next generation neural interfaces for humans. In this study, we evaluated the abiotic (electrophysiology, impedance, electrode morphology) and biotic (microglial reactivity, blood-brain barrier disruption, biochemical markers of axonal injury) effects of 16-channel, 50 μm diameter, polyimide insulated tungsten microwires array for implant durations that ranged from acute to up to 9 months in 25 rats. Daily electrode impedance spectroscopy, electrophysiological recordings, blood and cerebrospinal fluid (CSF) withdrawals, and histopathological analysis were performed to study the time-varying effects of chronic electrode implantation. Structural changes at the electrode recording site were observed as early as within 2-3 h of electrode insertion. Abiotic analysis indicated the first 2-3 weeks following surgery was the most dynamic period in the chronic electrode lifetime as there were greater variations in the electrode impedance, functional electrode performance, and the structural changes occurring at the electrode recording tips. Electrode recording site deterioration continued for the long-term chronic animals as insulation damage occurred and recording surface became more recessed over time. In general, electrode impedance and functional performance had smaller daily variations combined with reduced electrode recording site changes during the chronic phase. Histopathological studies were focused largely on characterizing microglial cell responses to electrode implantation. We found that activated microglia were present near the electrode tracks in all non-acute animals studied, thus indicating presence of a neuroinflammatory response regardless of post-implantation survival times and electrode performance. Conversely, dystrophic microglia detectable as fragmented cells were found almost exclusively in acute animals surviving only few hours after implantation. While there was no consistent relationship between microglial cell responses and electrode performance, we noticed co-occurrence of high ferritin expression, intraparenchymal bleeding, and microglial degeneration suggesting presence of excessive oxidative stress via Fenton chemistry. Biochemical analysis indicated that these electrodes always caused a persistent release of axonal injury biomarkers even several months after implantation suggesting persistent tissue damage. Our study suggests that mechanisms of electrode failure are multi-factorial involving both abiotic and biotic parameters. Since these failure modes occur concurrently and cannot be isolated from one another, the lack of consistent relationship between electrode performance and microglial responses in our results suggest that one or more of the abiotic factors were equally responsible for degradation in electrode performance over long periods of time.

UR - http://www.scopus.com/inward/record.url?scp=84866930516&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84866930516&partnerID=8YFLogxK

U2 - 10.1088/1741-2560/9/5/056015

DO - 10.1088/1741-2560/9/5/056015

M3 - Article

C2 - 23010756

AN - SCOPUS:84866930516

VL - 9

JO - Journal of Neural Engineering

JF - Journal of Neural Engineering

SN - 1741-2560

IS - 5

M1 - 056015

ER -