Clinical utility of a colorimetric end-tidal CO2 detector in cardiopulmonary resuscitation and emergency intubation

Albert J. Varon, John Morrina, Joseph M. Civetta

Research output: Contribution to journalArticle

58 Citations (Scopus)

Abstract

The purposes of this study were to evaluate the clinical utility of a colorimetric end-tidal CO2 (ETCO2) detector in confirming proper endotracheal intubation in patients requiring emergency intubation, to determine if this new device can be used as an adjunct to judge the effectiveness of cardiopulmonary resuscitation (CPR), and to determine whether the device can predict successful resuscitation from cardiopulmonary arrest. We studied prospectively 110 patients requiring emergency intubation for either respiratory distress (53 patients) or cardiopulmonary arrest (57 patients) by recording the color range of the indicator after the initial intubation. In patients who suffered a cardiopulmonary arrest, the color range was also recorded during CPR after the endotracheal tube was confirmed to be in the tracheal position and perfusion optimized, and at the moment CPR was stopped. The ETCO2 detector was 100% specific for correct endotracheal intubation in all patients. It was also highly sensitive (0.98) for correct endotracheal intubation in patients with respiratory distress. However, it was not sensitive (0.62) in patients with cardiopulmonary arrest and low perfusion. The sensitivity improved (0.88) when we used the ETCO2 range obtained after attempts to increase perfusion. A low ETCO2 color range in 19 patients undergoing CPR was interpreted as low cardiac output and prompted the physicians to attempt to increase perfusion. Of the patients who underwent CPR, no patient whose ETCO2 level remained less than 2% was successfully resuscitated. Those patients who had an ETCO2 level ≥2% had a significantly higher incidence of successful resuscitation. We conclude that the colorimetric ETCO2 detector is reliable and provides reassurance of correct endotracheal tube placement in patients requiring emergency intubation for respiratory distress. This device helps identify patients with low perfusion during CPR and is a useful prognostic indicator of successful short-term resuscitation.

Original languageEnglish
Pages (from-to)289-293
Number of pages5
JournalJournal of Clinical Monitoring
Volume7
Issue number4
DOIs
StatePublished - Oct 1 1991

Fingerprint

Cardiopulmonary Resuscitation
Intubation
Emergencies
Heart Arrest
Perfusion
Intratracheal Intubation
Resuscitation
Color
Equipment and Supplies
Low Cardiac Output

Keywords

  • Heart: cardiopulmonary resuscitation
  • Intubation: endotracheal, esophageal
  • Monitoring: carbon dioxide

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine

Cite this

Clinical utility of a colorimetric end-tidal CO2 detector in cardiopulmonary resuscitation and emergency intubation. / Varon, Albert J.; Morrina, John; Civetta, Joseph M.

In: Journal of Clinical Monitoring, Vol. 7, No. 4, 01.10.1991, p. 289-293.

Research output: Contribution to journalArticle

@article{de836a8b74834bb8ab2421bb9397fc3c,
title = "Clinical utility of a colorimetric end-tidal CO2 detector in cardiopulmonary resuscitation and emergency intubation",
abstract = "The purposes of this study were to evaluate the clinical utility of a colorimetric end-tidal CO2 (ETCO2) detector in confirming proper endotracheal intubation in patients requiring emergency intubation, to determine if this new device can be used as an adjunct to judge the effectiveness of cardiopulmonary resuscitation (CPR), and to determine whether the device can predict successful resuscitation from cardiopulmonary arrest. We studied prospectively 110 patients requiring emergency intubation for either respiratory distress (53 patients) or cardiopulmonary arrest (57 patients) by recording the color range of the indicator after the initial intubation. In patients who suffered a cardiopulmonary arrest, the color range was also recorded during CPR after the endotracheal tube was confirmed to be in the tracheal position and perfusion optimized, and at the moment CPR was stopped. The ETCO2 detector was 100{\%} specific for correct endotracheal intubation in all patients. It was also highly sensitive (0.98) for correct endotracheal intubation in patients with respiratory distress. However, it was not sensitive (0.62) in patients with cardiopulmonary arrest and low perfusion. The sensitivity improved (0.88) when we used the ETCO2 range obtained after attempts to increase perfusion. A low ETCO2 color range in 19 patients undergoing CPR was interpreted as low cardiac output and prompted the physicians to attempt to increase perfusion. Of the patients who underwent CPR, no patient whose ETCO2 level remained less than 2{\%} was successfully resuscitated. Those patients who had an ETCO2 level ≥2{\%} had a significantly higher incidence of successful resuscitation. We conclude that the colorimetric ETCO2 detector is reliable and provides reassurance of correct endotracheal tube placement in patients requiring emergency intubation for respiratory distress. This device helps identify patients with low perfusion during CPR and is a useful prognostic indicator of successful short-term resuscitation.",
keywords = "Heart: cardiopulmonary resuscitation, Intubation: endotracheal, esophageal, Monitoring: carbon dioxide",
author = "Varon, {Albert J.} and John Morrina and Civetta, {Joseph M.}",
year = "1991",
month = "10",
day = "1",
doi = "10.1007/BF01619347",
language = "English",
volume = "7",
pages = "289--293",
journal = "Journal of Clinical Monitoring and Computing",
issn = "1387-1307",
publisher = "Kluwer Academic Publishers",
number = "4",

}

TY - JOUR

T1 - Clinical utility of a colorimetric end-tidal CO2 detector in cardiopulmonary resuscitation and emergency intubation

AU - Varon, Albert J.

AU - Morrina, John

AU - Civetta, Joseph M.

PY - 1991/10/1

Y1 - 1991/10/1

N2 - The purposes of this study were to evaluate the clinical utility of a colorimetric end-tidal CO2 (ETCO2) detector in confirming proper endotracheal intubation in patients requiring emergency intubation, to determine if this new device can be used as an adjunct to judge the effectiveness of cardiopulmonary resuscitation (CPR), and to determine whether the device can predict successful resuscitation from cardiopulmonary arrest. We studied prospectively 110 patients requiring emergency intubation for either respiratory distress (53 patients) or cardiopulmonary arrest (57 patients) by recording the color range of the indicator after the initial intubation. In patients who suffered a cardiopulmonary arrest, the color range was also recorded during CPR after the endotracheal tube was confirmed to be in the tracheal position and perfusion optimized, and at the moment CPR was stopped. The ETCO2 detector was 100% specific for correct endotracheal intubation in all patients. It was also highly sensitive (0.98) for correct endotracheal intubation in patients with respiratory distress. However, it was not sensitive (0.62) in patients with cardiopulmonary arrest and low perfusion. The sensitivity improved (0.88) when we used the ETCO2 range obtained after attempts to increase perfusion. A low ETCO2 color range in 19 patients undergoing CPR was interpreted as low cardiac output and prompted the physicians to attempt to increase perfusion. Of the patients who underwent CPR, no patient whose ETCO2 level remained less than 2% was successfully resuscitated. Those patients who had an ETCO2 level ≥2% had a significantly higher incidence of successful resuscitation. We conclude that the colorimetric ETCO2 detector is reliable and provides reassurance of correct endotracheal tube placement in patients requiring emergency intubation for respiratory distress. This device helps identify patients with low perfusion during CPR and is a useful prognostic indicator of successful short-term resuscitation.

AB - The purposes of this study were to evaluate the clinical utility of a colorimetric end-tidal CO2 (ETCO2) detector in confirming proper endotracheal intubation in patients requiring emergency intubation, to determine if this new device can be used as an adjunct to judge the effectiveness of cardiopulmonary resuscitation (CPR), and to determine whether the device can predict successful resuscitation from cardiopulmonary arrest. We studied prospectively 110 patients requiring emergency intubation for either respiratory distress (53 patients) or cardiopulmonary arrest (57 patients) by recording the color range of the indicator after the initial intubation. In patients who suffered a cardiopulmonary arrest, the color range was also recorded during CPR after the endotracheal tube was confirmed to be in the tracheal position and perfusion optimized, and at the moment CPR was stopped. The ETCO2 detector was 100% specific for correct endotracheal intubation in all patients. It was also highly sensitive (0.98) for correct endotracheal intubation in patients with respiratory distress. However, it was not sensitive (0.62) in patients with cardiopulmonary arrest and low perfusion. The sensitivity improved (0.88) when we used the ETCO2 range obtained after attempts to increase perfusion. A low ETCO2 color range in 19 patients undergoing CPR was interpreted as low cardiac output and prompted the physicians to attempt to increase perfusion. Of the patients who underwent CPR, no patient whose ETCO2 level remained less than 2% was successfully resuscitated. Those patients who had an ETCO2 level ≥2% had a significantly higher incidence of successful resuscitation. We conclude that the colorimetric ETCO2 detector is reliable and provides reassurance of correct endotracheal tube placement in patients requiring emergency intubation for respiratory distress. This device helps identify patients with low perfusion during CPR and is a useful prognostic indicator of successful short-term resuscitation.

KW - Heart: cardiopulmonary resuscitation

KW - Intubation: endotracheal, esophageal

KW - Monitoring: carbon dioxide

UR - http://www.scopus.com/inward/record.url?scp=0026245219&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0026245219&partnerID=8YFLogxK

U2 - 10.1007/BF01619347

DO - 10.1007/BF01619347

M3 - Article

C2 - 1744672

AN - SCOPUS:0026245219

VL - 7

SP - 289

EP - 293

JO - Journal of Clinical Monitoring and Computing

JF - Journal of Clinical Monitoring and Computing

SN - 1387-1307

IS - 4

ER -