Approaches for optimizing the calibration standard of Tewameter TM 300

Mariya Miteva, Stefan Richter, Peter Elsner, Joachim W. Fluhr

Research output: Contribution to journalArticle

18 Citations (Scopus)

Abstract

Calibration of devices measuring transepidermal water loss (TEWL) is in intensive discussion. Comparative studies revealed that comparable measuring systems, e.g. open and closed chamber systems, do not always deliver the same results, even when expressing the measured values in SI units, namely in g/m2/h. Therefore, adequate and reliable calibration procedures need to be established. We were able to test the reliability of a multi-step calibration algorithm for an open chamber system such as Tewameter TM 300. In order to achieve reliable measurements, the maintenance of stable microclimate conditions without air turbulences is mandatory. The TEWL values should be compared with those determined gravimetrically on heated skin simulators. The reproducibility of the results is warranted by consecutive measurements on different adjacent spots of a defined area. Preheating of the probe sensors is an effective approach for shortening the measuring time and gaining a rapid steady-state. The accurate calibration of the probe can be checked under laboratory conditions any time. The critical point of the calibration and ultimately the accuracy of in vivo measurements maintain the steady functional capacity of the probes during the entire duration of continuous studies. The studied calibration procedure ensures these requirements.

Original languageEnglish (US)
Pages (from-to)904-912
Number of pages9
JournalExperimental Dermatology
Volume15
Issue number11
DOIs
StatePublished - Nov 1 2006
Externally publishedYes

Fingerprint

Calibration
International System of Units
Microclimate
Water
Preheating
Open systems
Reproducibility of Results
Skin
Turbulence
Simulators
Air
Maintenance
Equipment and Supplies
Sensors

Keywords

  • Air circulation
  • Ambient conditions
  • Barrier function
  • Calibration
  • Probe preheating
  • Skin simulator
  • Transepidermal water loss

ASJC Scopus subject areas

  • Biochemistry
  • Molecular Biology
  • Dermatology

Cite this

Approaches for optimizing the calibration standard of Tewameter TM 300. / Miteva, Mariya; Richter, Stefan; Elsner, Peter; Fluhr, Joachim W.

In: Experimental Dermatology, Vol. 15, No. 11, 01.11.2006, p. 904-912.

Research output: Contribution to journalArticle

Miteva, Mariya ; Richter, Stefan ; Elsner, Peter ; Fluhr, Joachim W. / Approaches for optimizing the calibration standard of Tewameter TM 300. In: Experimental Dermatology. 2006 ; Vol. 15, No. 11. pp. 904-912.
@article{37183982580a4422b97ef6543043ad2c,
title = "Approaches for optimizing the calibration standard of Tewameter TM 300",
abstract = "Calibration of devices measuring transepidermal water loss (TEWL) is in intensive discussion. Comparative studies revealed that comparable measuring systems, e.g. open and closed chamber systems, do not always deliver the same results, even when expressing the measured values in SI units, namely in g/m2/h. Therefore, adequate and reliable calibration procedures need to be established. We were able to test the reliability of a multi-step calibration algorithm for an open chamber system such as Tewameter TM 300. In order to achieve reliable measurements, the maintenance of stable microclimate conditions without air turbulences is mandatory. The TEWL values should be compared with those determined gravimetrically on heated skin simulators. The reproducibility of the results is warranted by consecutive measurements on different adjacent spots of a defined area. Preheating of the probe sensors is an effective approach for shortening the measuring time and gaining a rapid steady-state. The accurate calibration of the probe can be checked under laboratory conditions any time. The critical point of the calibration and ultimately the accuracy of in vivo measurements maintain the steady functional capacity of the probes during the entire duration of continuous studies. The studied calibration procedure ensures these requirements.",
keywords = "Air circulation, Ambient conditions, Barrier function, Calibration, Probe preheating, Skin simulator, Transepidermal water loss",
author = "Mariya Miteva and Stefan Richter and Peter Elsner and Fluhr, {Joachim W.}",
year = "2006",
month = "11",
day = "1",
doi = "10.1111/j.1600-0625.2006.00482.x",
language = "English (US)",
volume = "15",
pages = "904--912",
journal = "Experimental Dermatology",
issn = "0906-6705",
publisher = "Wiley-Blackwell",
number = "11",

}

TY - JOUR

T1 - Approaches for optimizing the calibration standard of Tewameter TM 300

AU - Miteva, Mariya

AU - Richter, Stefan

AU - Elsner, Peter

AU - Fluhr, Joachim W.

PY - 2006/11/1

Y1 - 2006/11/1

N2 - Calibration of devices measuring transepidermal water loss (TEWL) is in intensive discussion. Comparative studies revealed that comparable measuring systems, e.g. open and closed chamber systems, do not always deliver the same results, even when expressing the measured values in SI units, namely in g/m2/h. Therefore, adequate and reliable calibration procedures need to be established. We were able to test the reliability of a multi-step calibration algorithm for an open chamber system such as Tewameter TM 300. In order to achieve reliable measurements, the maintenance of stable microclimate conditions without air turbulences is mandatory. The TEWL values should be compared with those determined gravimetrically on heated skin simulators. The reproducibility of the results is warranted by consecutive measurements on different adjacent spots of a defined area. Preheating of the probe sensors is an effective approach for shortening the measuring time and gaining a rapid steady-state. The accurate calibration of the probe can be checked under laboratory conditions any time. The critical point of the calibration and ultimately the accuracy of in vivo measurements maintain the steady functional capacity of the probes during the entire duration of continuous studies. The studied calibration procedure ensures these requirements.

AB - Calibration of devices measuring transepidermal water loss (TEWL) is in intensive discussion. Comparative studies revealed that comparable measuring systems, e.g. open and closed chamber systems, do not always deliver the same results, even when expressing the measured values in SI units, namely in g/m2/h. Therefore, adequate and reliable calibration procedures need to be established. We were able to test the reliability of a multi-step calibration algorithm for an open chamber system such as Tewameter TM 300. In order to achieve reliable measurements, the maintenance of stable microclimate conditions without air turbulences is mandatory. The TEWL values should be compared with those determined gravimetrically on heated skin simulators. The reproducibility of the results is warranted by consecutive measurements on different adjacent spots of a defined area. Preheating of the probe sensors is an effective approach for shortening the measuring time and gaining a rapid steady-state. The accurate calibration of the probe can be checked under laboratory conditions any time. The critical point of the calibration and ultimately the accuracy of in vivo measurements maintain the steady functional capacity of the probes during the entire duration of continuous studies. The studied calibration procedure ensures these requirements.

KW - Air circulation

KW - Ambient conditions

KW - Barrier function

KW - Calibration

KW - Probe preheating

KW - Skin simulator

KW - Transepidermal water loss

UR - http://www.scopus.com/inward/record.url?scp=33749179054&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33749179054&partnerID=8YFLogxK

U2 - 10.1111/j.1600-0625.2006.00482.x

DO - 10.1111/j.1600-0625.2006.00482.x

M3 - Article

C2 - 17002688

AN - SCOPUS:33749179054

VL - 15

SP - 904

EP - 912

JO - Experimental Dermatology

JF - Experimental Dermatology

SN - 0906-6705

IS - 11

ER -