An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery

Cheng Yu, William Main, David Taylor, Gopinath Kuduvalli, Michael L J Apuzzo, John R. Adler, Michael Y. Wang

Research output: Contribution to journalArticle

106 Citations (Scopus)

Abstract

OBJECTIVE: Stereotactic radiosurgery requires the highest degree of accuracy in target identification and localization. When targeting paraspinal lesions, the CyberKnife radiosurgical system (Accuray, Inc., Sunnyvale, CA) uses implanted stainless steel fiducials. The purpose of this study was to evaluate the total system for clinically relevant accuracy of this approach. METHODS: The clinically relevant accuracy of the CyberKnife depends on 1) the accuracy of beam delivery, which in turn represents a compilation of robot and camera image-tracking errors, and 2) the inherent accuracy of target localization that stems from computed tomographic imaging and treatment planning. The clinically relevant accuracy was measured at three different CyberKnife facilities using head and torso phantoms loaded with packs of radiochromic film and expressed as a displacement of the dose contours from the treatment planning. RESULTS: The mean clinically relevant error, as measured at three different CyberKnife facilities, was determined to be 0.7 ± 0.3 mm, which did not vary with computed tomographic slice thickness in a range of 0.625 to 1.5 mm. The average treatment delivery precision was 0.3 ± 0.1 mm. Fiducial tracking error was less than 0.3 mm for radial translations up to 14 mm and less than 0.7 mm for rotations up to 4.5 degrees. CONCLUSION: For the treatment of relatively stationary spinal lesions targeted with fiducial tracking, the CyberKnife system is capable of submillimeter accuracy.

Original languageEnglish
Pages (from-to)1138-1146
Number of pages9
JournalNeurosurgery
Volume55
Issue number5
DOIs
StatePublished - Nov 1 2004
Externally publishedYes

Fingerprint

Radiosurgery
Torso
Stainless Steel
Therapeutics
Head

Keywords

  • Accuracy
  • CyberKnife
  • Spinal radiosurgery

ASJC Scopus subject areas

  • Clinical Neurology
  • Surgery

Cite this

Yu, C., Main, W., Taylor, D., Kuduvalli, G., Apuzzo, M. L. J., Adler, J. R., & Wang, M. Y. (2004). An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery. Neurosurgery, 55(5), 1138-1146. https://doi.org/10.1227/01.NEU.0000141080.54647.11

An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery. / Yu, Cheng; Main, William; Taylor, David; Kuduvalli, Gopinath; Apuzzo, Michael L J; Adler, John R.; Wang, Michael Y.

In: Neurosurgery, Vol. 55, No. 5, 01.11.2004, p. 1138-1146.

Research output: Contribution to journalArticle

Yu, C, Main, W, Taylor, D, Kuduvalli, G, Apuzzo, MLJ, Adler, JR & Wang, MY 2004, 'An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery', Neurosurgery, vol. 55, no. 5, pp. 1138-1146. https://doi.org/10.1227/01.NEU.0000141080.54647.11
Yu C, Main W, Taylor D, Kuduvalli G, Apuzzo MLJ, Adler JR et al. An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery. Neurosurgery. 2004 Nov 1;55(5):1138-1146. https://doi.org/10.1227/01.NEU.0000141080.54647.11
Yu, Cheng ; Main, William ; Taylor, David ; Kuduvalli, Gopinath ; Apuzzo, Michael L J ; Adler, John R. ; Wang, Michael Y. / An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery. In: Neurosurgery. 2004 ; Vol. 55, No. 5. pp. 1138-1146.
@article{97a620b1f3c848afac5c1d4a59c59432,
title = "An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery",
abstract = "OBJECTIVE: Stereotactic radiosurgery requires the highest degree of accuracy in target identification and localization. When targeting paraspinal lesions, the CyberKnife radiosurgical system (Accuray, Inc., Sunnyvale, CA) uses implanted stainless steel fiducials. The purpose of this study was to evaluate the total system for clinically relevant accuracy of this approach. METHODS: The clinically relevant accuracy of the CyberKnife depends on 1) the accuracy of beam delivery, which in turn represents a compilation of robot and camera image-tracking errors, and 2) the inherent accuracy of target localization that stems from computed tomographic imaging and treatment planning. The clinically relevant accuracy was measured at three different CyberKnife facilities using head and torso phantoms loaded with packs of radiochromic film and expressed as a displacement of the dose contours from the treatment planning. RESULTS: The mean clinically relevant error, as measured at three different CyberKnife facilities, was determined to be 0.7 ± 0.3 mm, which did not vary with computed tomographic slice thickness in a range of 0.625 to 1.5 mm. The average treatment delivery precision was 0.3 ± 0.1 mm. Fiducial tracking error was less than 0.3 mm for radial translations up to 14 mm and less than 0.7 mm for rotations up to 4.5 degrees. CONCLUSION: For the treatment of relatively stationary spinal lesions targeted with fiducial tracking, the CyberKnife system is capable of submillimeter accuracy.",
keywords = "Accuracy, CyberKnife, Spinal radiosurgery",
author = "Cheng Yu and William Main and David Taylor and Gopinath Kuduvalli and Apuzzo, {Michael L J} and Adler, {John R.} and Wang, {Michael Y.}",
year = "2004",
month = "11",
day = "1",
doi = "10.1227/01.NEU.0000141080.54647.11",
language = "English",
volume = "55",
pages = "1138--1146",
journal = "Neurosurgery",
issn = "0148-396X",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery

AU - Yu, Cheng

AU - Main, William

AU - Taylor, David

AU - Kuduvalli, Gopinath

AU - Apuzzo, Michael L J

AU - Adler, John R.

AU - Wang, Michael Y.

PY - 2004/11/1

Y1 - 2004/11/1

N2 - OBJECTIVE: Stereotactic radiosurgery requires the highest degree of accuracy in target identification and localization. When targeting paraspinal lesions, the CyberKnife radiosurgical system (Accuray, Inc., Sunnyvale, CA) uses implanted stainless steel fiducials. The purpose of this study was to evaluate the total system for clinically relevant accuracy of this approach. METHODS: The clinically relevant accuracy of the CyberKnife depends on 1) the accuracy of beam delivery, which in turn represents a compilation of robot and camera image-tracking errors, and 2) the inherent accuracy of target localization that stems from computed tomographic imaging and treatment planning. The clinically relevant accuracy was measured at three different CyberKnife facilities using head and torso phantoms loaded with packs of radiochromic film and expressed as a displacement of the dose contours from the treatment planning. RESULTS: The mean clinically relevant error, as measured at three different CyberKnife facilities, was determined to be 0.7 ± 0.3 mm, which did not vary with computed tomographic slice thickness in a range of 0.625 to 1.5 mm. The average treatment delivery precision was 0.3 ± 0.1 mm. Fiducial tracking error was less than 0.3 mm for radial translations up to 14 mm and less than 0.7 mm for rotations up to 4.5 degrees. CONCLUSION: For the treatment of relatively stationary spinal lesions targeted with fiducial tracking, the CyberKnife system is capable of submillimeter accuracy.

AB - OBJECTIVE: Stereotactic radiosurgery requires the highest degree of accuracy in target identification and localization. When targeting paraspinal lesions, the CyberKnife radiosurgical system (Accuray, Inc., Sunnyvale, CA) uses implanted stainless steel fiducials. The purpose of this study was to evaluate the total system for clinically relevant accuracy of this approach. METHODS: The clinically relevant accuracy of the CyberKnife depends on 1) the accuracy of beam delivery, which in turn represents a compilation of robot and camera image-tracking errors, and 2) the inherent accuracy of target localization that stems from computed tomographic imaging and treatment planning. The clinically relevant accuracy was measured at three different CyberKnife facilities using head and torso phantoms loaded with packs of radiochromic film and expressed as a displacement of the dose contours from the treatment planning. RESULTS: The mean clinically relevant error, as measured at three different CyberKnife facilities, was determined to be 0.7 ± 0.3 mm, which did not vary with computed tomographic slice thickness in a range of 0.625 to 1.5 mm. The average treatment delivery precision was 0.3 ± 0.1 mm. Fiducial tracking error was less than 0.3 mm for radial translations up to 14 mm and less than 0.7 mm for rotations up to 4.5 degrees. CONCLUSION: For the treatment of relatively stationary spinal lesions targeted with fiducial tracking, the CyberKnife system is capable of submillimeter accuracy.

KW - Accuracy

KW - CyberKnife

KW - Spinal radiosurgery

UR - http://www.scopus.com/inward/record.url?scp=8744294785&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=8744294785&partnerID=8YFLogxK

U2 - 10.1227/01.NEU.0000141080.54647.11

DO - 10.1227/01.NEU.0000141080.54647.11

M3 - Article

VL - 55

SP - 1138

EP - 1146

JO - Neurosurgery

JF - Neurosurgery

SN - 0148-396X

IS - 5

ER -