A novel membrane sensor for histamine H1-receptor antagonist "fexofenadine"

M. N. Abbas, A. A.Abdel Fattah, Elsayed Zahran

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

The construction and general performance of thirteen new polymeric membrane sensors for the determination of fexofenadine hydrochloride based on its ion exchange with reineckate, tetraphenylborate and tetraiodomercurate have been studied. The effects of membrane composition, type of plasticizer, pH value of sample solution and concentration of the analyte in the sensor internal solution have been thoroughly investigated. The novel sensor based on reineckate exchanger shows a stable, potentiometric response for fexofenadine in the concentration range of 1 × 10-2 - 2.5 × 10-6 M at 25°C that is independent of pH in the range of 2.0 - 4.5. The sensor possesses a Nemstian cationic slope of 62.3 ± 0.7 mV/concentration decade and a lower detection limit of 1.3 × 10-6 M with a fast response time of 20 - 40 s. Selectivity coefficients for a number of interfering ions and excipients relative to fexofenadine were investigated. There is negligible interference from almost all studied cations, anions, and pharmaceutical excipients, however, citrizine that has a structure homologous to that of fexofenadine was found to interfere. The determination of fexofenadine in aqueous solution shows an average recovery of 99.83% with a mean relative standard deviation (RSD) of 0.5%. Direct potentiometric determination of fexofenadine in tablets gave results that compare favorably with those obtained by standard spectrophotometric methods. Potentiometric titration of fexofenadine with phosphomolybdic acid as a titrant has been monitored with the proposed sensor as an end point indicator electrode.

Original languageEnglish (US)
Pages (from-to)1137-1142
Number of pages6
JournalAnalytical Sciences
Volume20
Issue number8
DOIs
StatePublished - Aug 1 2004
Externally publishedYes

Fingerprint

fexofenadine
Histamine H1 Antagonists
Membranes
Sensors
Excipients
Tetraphenylborate
Plasticizers
Polymeric membranes
Titration
Tablets
Anions
Cations

ASJC Scopus subject areas

  • Analytical Chemistry

Cite this

A novel membrane sensor for histamine H1-receptor antagonist "fexofenadine". / Abbas, M. N.; Fattah, A. A.Abdel; Zahran, Elsayed.

In: Analytical Sciences, Vol. 20, No. 8, 01.08.2004, p. 1137-1142.

Research output: Contribution to journalArticle

Abbas, M. N. ; Fattah, A. A.Abdel ; Zahran, Elsayed. / A novel membrane sensor for histamine H1-receptor antagonist "fexofenadine". In: Analytical Sciences. 2004 ; Vol. 20, No. 8. pp. 1137-1142.
@article{bfe8e167b8ef49af9cb91de74326428b,
title = "A novel membrane sensor for histamine H1-receptor antagonist {"}fexofenadine{"}",
abstract = "The construction and general performance of thirteen new polymeric membrane sensors for the determination of fexofenadine hydrochloride based on its ion exchange with reineckate, tetraphenylborate and tetraiodomercurate have been studied. The effects of membrane composition, type of plasticizer, pH value of sample solution and concentration of the analyte in the sensor internal solution have been thoroughly investigated. The novel sensor based on reineckate exchanger shows a stable, potentiometric response for fexofenadine in the concentration range of 1 × 10-2 - 2.5 × 10-6 M at 25°C that is independent of pH in the range of 2.0 - 4.5. The sensor possesses a Nemstian cationic slope of 62.3 ± 0.7 mV/concentration decade and a lower detection limit of 1.3 × 10-6 M with a fast response time of 20 - 40 s. Selectivity coefficients for a number of interfering ions and excipients relative to fexofenadine were investigated. There is negligible interference from almost all studied cations, anions, and pharmaceutical excipients, however, citrizine that has a structure homologous to that of fexofenadine was found to interfere. The determination of fexofenadine in aqueous solution shows an average recovery of 99.83{\%} with a mean relative standard deviation (RSD) of 0.5{\%}. Direct potentiometric determination of fexofenadine in tablets gave results that compare favorably with those obtained by standard spectrophotometric methods. Potentiometric titration of fexofenadine with phosphomolybdic acid as a titrant has been monitored with the proposed sensor as an end point indicator electrode.",
author = "Abbas, {M. N.} and Fattah, {A. A.Abdel} and Elsayed Zahran",
year = "2004",
month = "8",
day = "1",
doi = "10.2116/analsci.20.1137",
language = "English (US)",
volume = "20",
pages = "1137--1142",
journal = "Analytical Sciences",
issn = "0910-6340",
publisher = "Japan Society for Analytical Chemistry",
number = "8",

}

TY - JOUR

T1 - A novel membrane sensor for histamine H1-receptor antagonist "fexofenadine"

AU - Abbas, M. N.

AU - Fattah, A. A.Abdel

AU - Zahran, Elsayed

PY - 2004/8/1

Y1 - 2004/8/1

N2 - The construction and general performance of thirteen new polymeric membrane sensors for the determination of fexofenadine hydrochloride based on its ion exchange with reineckate, tetraphenylborate and tetraiodomercurate have been studied. The effects of membrane composition, type of plasticizer, pH value of sample solution and concentration of the analyte in the sensor internal solution have been thoroughly investigated. The novel sensor based on reineckate exchanger shows a stable, potentiometric response for fexofenadine in the concentration range of 1 × 10-2 - 2.5 × 10-6 M at 25°C that is independent of pH in the range of 2.0 - 4.5. The sensor possesses a Nemstian cationic slope of 62.3 ± 0.7 mV/concentration decade and a lower detection limit of 1.3 × 10-6 M with a fast response time of 20 - 40 s. Selectivity coefficients for a number of interfering ions and excipients relative to fexofenadine were investigated. There is negligible interference from almost all studied cations, anions, and pharmaceutical excipients, however, citrizine that has a structure homologous to that of fexofenadine was found to interfere. The determination of fexofenadine in aqueous solution shows an average recovery of 99.83% with a mean relative standard deviation (RSD) of 0.5%. Direct potentiometric determination of fexofenadine in tablets gave results that compare favorably with those obtained by standard spectrophotometric methods. Potentiometric titration of fexofenadine with phosphomolybdic acid as a titrant has been monitored with the proposed sensor as an end point indicator electrode.

AB - The construction and general performance of thirteen new polymeric membrane sensors for the determination of fexofenadine hydrochloride based on its ion exchange with reineckate, tetraphenylborate and tetraiodomercurate have been studied. The effects of membrane composition, type of plasticizer, pH value of sample solution and concentration of the analyte in the sensor internal solution have been thoroughly investigated. The novel sensor based on reineckate exchanger shows a stable, potentiometric response for fexofenadine in the concentration range of 1 × 10-2 - 2.5 × 10-6 M at 25°C that is independent of pH in the range of 2.0 - 4.5. The sensor possesses a Nemstian cationic slope of 62.3 ± 0.7 mV/concentration decade and a lower detection limit of 1.3 × 10-6 M with a fast response time of 20 - 40 s. Selectivity coefficients for a number of interfering ions and excipients relative to fexofenadine were investigated. There is negligible interference from almost all studied cations, anions, and pharmaceutical excipients, however, citrizine that has a structure homologous to that of fexofenadine was found to interfere. The determination of fexofenadine in aqueous solution shows an average recovery of 99.83% with a mean relative standard deviation (RSD) of 0.5%. Direct potentiometric determination of fexofenadine in tablets gave results that compare favorably with those obtained by standard spectrophotometric methods. Potentiometric titration of fexofenadine with phosphomolybdic acid as a titrant has been monitored with the proposed sensor as an end point indicator electrode.

UR - http://www.scopus.com/inward/record.url?scp=4344665733&partnerID=8YFLogxK